Suzhou Macxi Fluid Technology Co.,Ltd.
Huis> Blog> BT-100D peristaltic pump|constant flow pump instruction manual

BT-100D peristaltic pump|constant flow pump instruction manual

July 19, 2023
BT-100D|BT-200D timing computer subdivision constant current pump instruction manual I, overview
BT-100D BT-200D constant current pump is a new type of timing computer subdivision driven peristaltic pump developed by our factory. It is a new generation of BT series pump. In addition to providing the speed control (0.1RPM-150RPM) of the older generation products, this series of products has newly added timing control: timing speed, timing sleep, external trigger and other air energy control.
The pump has the characteristics of low operating noise, simple operation, stable performance, power-off data storage, large flow, and timing. It can be widely used in environmental protection, medicine, food, chemical and other fields, as well as universities and colleges and other laboratories. related industry.
Second, the main technical parameters
1, flow BT-100SD 10-14500ML/H
BT-200SD 2×(10-14500ML/H) is a dual pump head dual channel type
2, pressure ≥ 2Kg
3, timing operation: 1 second - 9 hours 59 minutes 59 seconds
4, timed sleep: 0 seconds - 9 hours 59 minutes 59 seconds
5, power requirements: AC220V ± 10% 50HZ 10W
6, the use of the environment: temperature 0-40 ° C humidity: ≤ 85%
7, continuous working time: more than 200 hours
8, flow accuracy: ± 2%
9, weight; 3KG
10, Dimensions: 290 × 110 × 155mm
Third, the button function introduction
1, ▲ button in the machine running and pause state, click this button to +1 incremental speed, press and hold to quickly increase the running speed.
2, ▼ In the machine running and pause state, click this button to -1 reduce the speed of the operation, press and hold to quickly reduce the running speed.
3. When the “Function” button is in the pause state, press this button to change the working mode of the pump. In the “normal mode”, “Timed operation setting”, “Timed sleep” setting, the number of cycles is set incorrectly. In the Try Port Selection dialog box, select another serial port or check the PC serial port parameters (see "PC and Device Section").
2) Equipment control:
The actual operation is the same as the panel operation. Please refer to the “Key Function Introduction and Mode Description” section above.
3) Device disconnected:
● Click on "Menu" -> "Device" -> "Disconnect".
● Click "OK" in the demand confirmation dialog.
● At this time, the system title shows “Device not connected” and the status shows the status before the connection is restored. The decimal point of the first digital tube on the left side of the device panel is off.
six. Instructions:
1. When loading the pump head, first move the moving rod to raise the pressure block, place the required hose on the roller (to straighten the hose outward), and then lower the pressure block to tighten the hose. . Put the two ends of the hose into the media of the output and input containers to start working. When not working, lift the clamp to loosen the pressure tube and extend the service life of the hose.
2, turn on the power switch, the digital tube displays "P" and the speed of the last operation, the pump waits for the start, press the "◢╱║" button, the pump runs according to the speed and direction stored in the microcomputer last time, the digital tube display The current speed of the pump.
3. When the peristaltic pump is running, press and hold the “◢◢╱◢” button to maximize the instantaneous speed of the pump, remove the residual gas in the silicone tube, release the button, and return the pump to its original speed. Press the “▲” or “▼” button to increase or decrease the pump speed. Press and hold the “▲” and “▼” buttons to make the pump speed rise or fall rapidly. Press the “◢◢╱◢” button to change the current running direction of the pump, from pumping to infusion, or changing the infusion to pumping.
4. After starting up, press the “◢╱║” button to change the pump from non-operating state to running state. While the pump is running, press the “◢╱║” button to pause the pump and save the running status (speed, direction) in the microcomputer. The digital tube displays “P---” and waits for the power to be turned off. If you do not need to save the data, you can close the power switch and end the operation.
Seven. How to convert the speed to flow:
Since the flow rate of the peristaltic pump is related to various factors such as the rotational speed, the degree of compression of the silicone tube, and the nature of the liquid, the flow rate originally set must be re-corrected for each operation to achieve constant current control. Here, a simple speed is introduced. The method of transforming to traffic.
1. Turn on the power switch and press the “◢╱║” button. After the pump is running, press and hold the “Exhaust” button to drain the residual gas in the silicone tube. Select a certain speed (for example: 42 rpm / M), and then press the "◢╱║" button, the pump stops running to save the speed at this time, turn off the power switch.
2. Put the silicone tube into the measuring cup, turn on the power switch, press the “◢╱║” button, and time the battery. When the pump runs to a certain time (such as 10 minutes), turn off the power switch and stop the pump.
3. Divide the amount of liquid in the measuring cup by the counting time to get the actual flow rate of the operation. If the amount of liquid in the measuring cup is 26 ml/10 M = 2.6 ml / 1 M, the current speed of 42 rpm / M is equivalent to a flow rate of 2.6 ml / M, and other required flow rates can be obtained in a similar manner.
Eight. Precautions for use:
1. During the use of the instrument, it should be noted that the silicone tube is in good condition at all times. It is generally better to operate at medium speed. When the silicone tube is found to be aging, it should be replaced in time. When the instrument starts to run, the silicone tube may have a swelling at one end, which can be straightened.
2. If it is not used for a long time, remove the silicone tube and extend the service period.
3, the instrument should be carefully kept, should be placed in a dry, ventilated, non-corrosive gas.
4. If the instrument is damaged within one year from the factory, all the manufacturer's manufacturing quality problems will be repaired free of charge by the factory.
nine. annex:
1, the manual
2. Certificate of conformity
3. Warranty card
4, with the collector connection line
5, RS232 cable
6, the power cord
7, fuse (1A) two
8, silicone tube (φ3.1X6.3φ4.8X8.0) each 2 meters
Neem contact op

Author:

Mr. Bruce Tang

Phone/WhatsApp:

+8613773149257

populaire producten
You may also like
Related Categories

E-mail naar dit bedrijf

Onderwerp:
E-mail:
bericht:

Your message must be betwwen 20-8000 characters

Neem contact op

Author:

Mr. Bruce Tang

Phone/WhatsApp:

+8613773149257

populaire producten

Huis

Product

Whatsapp

Over ons

onderzoek

We will contact you immediately

Fill in more information so that we can get in touch with you faster

Privacy statement: Your privacy is very important to Us. Our company promises not to disclose your personal information to any external company with out your explicit permission.

verzenden